Category: Chromatography - Liquid/HPLC | Subcategory: Autosamplers
Agilent 1100 Series G1330A ALS Therm

ITEM: 18396 In Stock




The Agilent 1100 Series G1330A ALS Therm is a temperature control unit for the G1329 Autosampler.
The G1330A ALS Therm controls temperature for either HPLCs or fraction collectors. The ALS thermostat contains Peltier-controlled heat exchangers. A fan draws air from the area above the sample vial tray of the autosampler and is then blown through the fins of the cooling/heating module. There it is cooled or heated according to the temperature setting. The thermostatted air enters the autosampler through a recess underneath the specially-designed sample tray. The air is then distributed evenly through the sample tray ensuring effective temperature control, regardless of how many vials are in the tray.
Specifications
- Operating Temperature Range: 4°C to 40°C
- Operating Humidity Range: < 95% non-condensing
- Max Power Consumption: 260 VA (210 W)
- BTU: 717