Your Trusted Source for Used Lab Equipment
Category: HPLC, UPLC, & LC/MS | Subcategory: HPLC Systems | Manufacturer: Agilent
Agilent 1200 Series HPLC System With G1311 Quat Pump, G1367 HiP ALS, G1330 ALS Therm, G1365 MWD
ITEM: 33782 In Stock
ALT Pay: $333.39/mo
Your Trusted Source for Used Lab Equipment
Free 90 Day Warranty
Up to 3 Year Warranty Upgrade Available
Tested & Refurbished Prior to Shipping
The Agilent 1200 Series HPLC system is a high-throughput chromatographic solution designed to deliver accurate gradient formation, rapid injection cycles, and sensitive UV-Vis detection. This configuration includes the G1311A quaternary pump for flexible solvent mixing, the G1367B high-performance autosampler with G1330B thermostat for cooled operation, and the G1365 multi-wavelength detector for simultaneous multi-channel detection—ideal for analytical labs requiring speed, precision, and flexibility.
- Quaternary pump allows low-pressure mixing of up to four solvents with excellent compositional precision
- High-performance autosampler delivers rapid injections and supports 2 mL vials or well plates
- Thermostatted autosampler keeps samples stable down to 4°C via G1330B
- MWD provides real-time detection at up to 8 wavelengths from 190 to 950 nm
Specifications:
- G1311A Quaternary Pump:
- Flow rate: 0.2 – 10.0 mL/min; settable in 0.001 mL/min increments
- Pressure range: up to 400 bar (5880 psi) at ≤5 mL/min
- Flow accuracy: ±1% or 10 µL/min
- Gradient formation: Low-pressure quaternary mixing; delay volume 800–1100 µL
- Power: 220 VA; Voltage: 100–240 VAC, 50/60 Hz
- Dimensions: 5.5 x 13.5 x 17 in (140 x 345 x 435 mm); Weight: 25 lbs (11 kg)
- G1367B High-Performance Autosampler with G1330B Thermostat:
- Injection range: 0.1–100 µL; up to 1500 µL with hardware modification
- Cycle time: Typically <30 seconds per injection
- Carryover: Typically <0.01%
- Sample formats: 2 mL vials, well plates, Eppendorf tubes
- Cooling range: Down to 4°C via G1330B thermostat
- Power: 300 VA max
- Dimensions: 8 x 13.5 x 17 in (200 x 345 x 435 mm); Weight: 34.2 lbs (15.5 kg)
- G1365 Multiple Wavelength Detector:
- Wavelength range: 190–950 nm
- Simultaneous detection: Up to 8 wavelengths
- Slit width: Programmable (1, 2, 4, 8, 16 nm)
- Flow cells: Standard 13 µL, 10 mm path length, up to 120 bar
- Noise: ±1 × 10−5 AU at 254/750 nm
- Wavelength accuracy: ±1 nm; Self-calibrating
- Power: 300 VA / 125 W
- Dimensions: 5.5 x 13.5 x 17 in (140 x 345 x 435 mm); Weight: 26 lbs (11.5 kg)
This system delivers fast, accurate injections, programmable multi-solvent gradients, and high-sensitivity UV-Vis detection—ideal for complex workflows and regulated lab environments.