The Agilent 1200 Series HPLC System with G1314B Variable Wavelength Detector, G1311A Quaternary Pump, and G1329A Autosampler is an integrated, modular liquid chromatography platform engineered for high precision and reliability in analytical applications. Designed for scalability and performance, this system is ideal for both routine and advanced separation workflows in pharmaceutical, environmental, food, and academic laboratories.
Key Features & Benefits:
- G1311A Quaternary Pump: Offers a wide flow range (0.001–10 mL/min), high flow precision (≤0.07% RSD), and low-pressure quaternary mixing for flexible solvent combinations and accurate gradient formation. Operating pressure reaches up to 5880 psi (40 MPa).
- G1329A Autosampler: Supports injection volumes from 0.1 to 900 µL with a precision of <0.25% RSD (for 5–100 µL) and sample capacity for up to 100 × 2 mL vials. Compatible with viscous samples (0.2–50 cP) and features low carryover (<0.1%).
- G1314B Variable Wavelength Detector: Covers a wavelength range of 190–600 nm, featuring a deuterium lamp source, ±1 nm wavelength accuracy, and various flow cell options. Linearity exceeds 2 AU and noise levels are low (±0.75 × 10−5 AU at 254 nm) for high signal clarity.
- Flexible Communication & Control: Integrated via CAN, GPIB, RS-232, or optional LAN, and operated through Agilent ChemStation for LC for streamlined data acquisition and method development.
Specifications:
- Pump Flow Range: 0.001 – 10 mL/min
- Pump Pressure Range: Up to 5880 psi (400 bar)
- Autosampler Injection Volume: 0.1 – 900 µL in 0.1 µL increments
- Detector Wavelength Range: 190 – 600 nm
- Module Dimensions (each): 5.5″ × 13.5″ × 17″ (140 × 345 × 435 mm)
- Electrical: 100–240 V, 50/60 Hz
With precision pumping, versatile injection, and sensitive detection, the Agilent 1200 Series HPLC System offers a comprehensive and scalable solution for accurate and reproducible chromatographic analyses.