Category: Chromatography - Liquid/HPLC | Subcategory: Detectors

Agilent 1260 Infinity II DAD WR (G7115A)

Agilent 1260 Infinity II DAD WR (G7115A) Image

ITEM: 32373 In Stock

  • Free 90 Day Warranty
  • One Year Upgrade Available
  • Tested & Serviced Prior to Shipping

Or call (860) 691-2213

Agilent 1260 Infinity II DAD WR (G7115A) ImageAgilent 1260 Infinity II DAD WR (G7115A) ImageAgilent 1260 Infinity II DAD WR (G7115A) ImageAgilent 1260 Infinity II DAD WR (G7115A) Image

The Agilent 1260 Infinity II DAD WR (G7115A) offers multiple wavelength and full spectral detection up to a 120 Hz sampling rate. Maximum flexibility is guaranteed with a wide wavelength range of 190 to 950 nm with dual-lamp design, as well as the choice of 15 analytical, preparative, and SFC flow cells. Even at high data rates, the Agilent 1260 Infinity II DAD WR (G7115A) enables precise identification, quantification, and peak purity analysis at trace levels.

Specifications:

Detection type: 1024-element photodiode array
Light source: Deuterium and tungsten lamps. 
Data rate: up to 120 Hz
Wavelength range: 190 – 950 nm
Short term noise (ASTM) Single and Multi-Wavelength: < ± 0.7·10-5 AU at 254 and 750 nm
Drift: < 0.9·10-3 AU/h at 254 nm
Linear absorbance range: > 2 AU (5 %) at 265 nm
Wavelength accuracy: ± 1 nm; Self-calibration with deuterium lines, verification with holmium oxide filter
Wavelength bunching: 1 – 400 nm Programmable in steps of 1 nm
Slit width: 1, 2, 4 , 8, 16 nm Programmable slit
Diode width: < 1 nm
Power: 100-240V, 60 Hz
Weight: 26.5 lbs (12 kg) 
Dimensions: 15.6" W x 17" D x 5.5" H inches (39.6 x 43.6 14 cm)

The ALT Difference

All Items Tested to
Manufacturer Spec

More Than 12,000
Items in Stock

Up to 1-Year
Warranty

Price Match
Guarantee

50%–80%
off Retail