Category: Incubators - Ovens | Subcategory: Environmental Chambers
Binder Series KBF 1020 - Humidity Test Chamber
ITEM: 25703
This product is currently out of stock, please check our complete line of in-stock Environmental Chambers by clicking the green link below or give one of the members of our sales team a call today at (860) 691-2213 so we can help you find a suitable replacement.
NEW from manufacturer! Comes with 2 year manufacturer's warranty.
Binder Series KBF 1020 - Humidity Test Chamber
Designation | Series KBF |
Series feature | Constant climate chambers with large temperature / humidity range |
Important characteristics | * Temperature range: 0 °C to 70 °C * Humidity range: 10 % to 80 % RH * APT.line™ preheating chamber technology * Humidity regulation with capacitive humidity sensor and vapor humidification * Inner chamber made of stainless steel * Intuitive touchscreen controller with time-segment and real-time programming with internal data recording * Tight-sealing inner door made of safety glass (ESG) with special TIMELESS coating * 2 stainless steel racks * Access port with silicone plug, 30 mm, left * 4 stable castors, two with brakes, from 240 liters * Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm * Computer interface: Ethernet * Door heating |
Features | * Temperature range: 0 °C to 70 °C * Humidity range: 10 % to 80 % RH * Humidity regulation with capacitive humidity sensor and vapor humidification * Intuitive touchscreen controller with time-segment and real-time programming with internal data recording * Inner chamber made of stainless steel |
Text of Series | The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climate conditions. From programming to documentation, this constant climate chamber meets all critical requirements. |
TECHNICAL DATA
Description | KBF1020-230V * | KBF1020-240V * |
Article Number | 9020-0326 | 9020-0327 |
Performance Data Temperature | ||
Temperature range | 0…70 | 0…70 |
Temperature variation at 40 °C [± K] | 0.2 | 0.2 |
Temperature fluctuation | 0,5 ± K | 0,5 ± K |
Max. heat compensation at 40 °C [W] | 600 | 600 |
Performance Data Climate | ||
Temperature range [°C] | 10…70 | 10…70 |
Temperature variation at 25 °C and 60 % RH [± K] | 0.2 | 0.2 |
Temperature variation at 40 °C and 75 % RH [± K] | 0.2 | 0.2 |
Temperature fluctuation at 25 °C and 60 % RH [± K] | 0.1 | 0.1 |
Temperature fluctuation at 40 °C and 75 % RH [± K] | 0.1 | 0.1 |
Humidity range | 10…80 | 10…80 |
Humidity fluctuation at 25 °C and 60 % RH | 1,5 ± % RH | 1,5 ± % RH |
Humidity fluctuation at 40 °C and 75 % RH | 1,5 ± % RH | 1,5 ± % RH |
Electrical data | ||
Rated Voltage [V] | 200…230 | 200…240 |
Power frequency [Hz] | 50/60 | 50/60 |
Nominal power [kW] | 3.1 | 3.1 |
Unit fuse [A] | 16 | 16 |
Phase (Nominal voltage) [ph] | 1 | 1 |
Measures - Outer dimensions | ||
Width net [mm] | 1250 | 1250 |
Height net [mm] | 1925 | 1925 |
Depth net [mm] | 1145 | 1145 |
Wall clearance back [mm] | 100 | 100 |
Wall clearance sidewise [mm] | 100 | 100 |
Measures - Doors | ||
Inner doors | 2 | 2 |
Unit doors | 2 | 2 |
Measures - Internal Dimensions | ||
Width [mm] | 976 | 976 |
Height [mm] | 1250 | 1250 |
Depth [mm] | 836 | 836 |
Measures | ||
Interior volume [L] | 1020 | 1020 |
Net weight of the unit (empty) [kg] | 365 | 365 |
permitted load [kg] | 150 | 150 |
Load per rack [kg] | 45 | 45 |
Environment-specific data | ||
Energy consumption at 40 °C and 75 % RH [Wh/h] | 650 | 650 |
Sound-pressure level [dB(A)] | 56 | 56 |
Fixtures | ||
Number of shelves (std./max.) | 2/15 | 2/15 |
* All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.