Category: Refrigerators - Freezers | Subcategory: Lab Freezers
Binder Series UF V 500 - Ultralow Temperature Freezer
ITEM: 25670
This product is currently out of stock, please check our complete line of in-stock Lab Freezers by clicking the green link below or give one of the members of our sales team a call today at (860) 691-2213 so we can help you find a suitable replacement.
NEW from manufacturer! Comes with 2 year manufacturer's warranty.
Binder Series UF V 500 - Ultralow Temperature Freezer
Designation | Series UF V |
Series feature | Ultralow temperature freezers with VIP insulation and electromechanical door lock |
Important characteristics | * Temperature range: -40 °C to -86 °C * Cooling with cascade compressor cooling unit * CFC-free safety refrigerants R-404A and R-508B * Door hinged right or left * RFID controlled Access (depending on version) * Data Logger USB (depending on version) * Water cooling (depending on version) * Electromechanical door latch * 4 compartment doors made of ABS plastic * Inner chamber made of stainless steel * Thermal insulation using vacuum insulating panels and PUR foam * Battery-backed alarm system * Computer interface: RS 422 * Analog output, 4 – 20 mA * Zero-voltage alarm contact * Elapsed time indicator * 3 stainless steel shelves |
Features | * Temperature range: -40 °C to -86 °C * Thermal insulation using vacuum insulating panels and PUR foam * Electromechanical door latch * RFID controlled Access (depending on version) * Data Logger USB (depending on version) * Water cooling (depending on version) * Battery-backed alarm system * Inner chamber made of stainless steel |
Text of Series | The BINDER ultralow temperature freezer ensures safe long-term storage of samples at a temperature of -86 °C. It is based on a multi-stage security concept, which offers everything from reliable operation to ease of use and system integration. |
TECHNICAL DATA
Description | UFV500-230V-R * | UFV500-230V-L * | UFV500-230V-RW * | UFV500-230V-LW * |
Article Number | 9020-0236 | 9020-0213 | 9020-0259 | 9020-0260 |
Performance Data Temperature | ||||
Temperature range | -86…-40 | -86…-40 | -86…-40 | -86…-40 |
Temperature variation at -80 °C [± K] | 2.5 | 2.5 | 2.5 | 2.5 |
Cooling down time from 25 °C to -80 °C [min] | 360 | 360 | 360 | 360 |
Warm-up time during power failure from -80 °C / -122 °F to -60 °C / -76 °F [min] | 150 | 150 | 150 | 150 |
Electrical data | ||||
Rated Voltage [V] | 230 | 230 | 230 | 230 |
Power frequency [Hz] | 50 | 50 | 50 | 50 |
Nominal power [kW] | 1.1 | 1.1 | 1.1 | 1.1 |
Unit fuse [A] | 10 | 10 | 10 | 10 |
Phase (Nominal voltage) [ph] | 1 | 1 | 1 | 1 |
Measures - Outer dimensions | ||||
Width net [mm] | 830 | 830 | 830 | 830 |
Height net [mm] | 1985 | 1985 | 1985 | 1985 |
Depth net [mm] | 885 | 885 | 885 | 885 |
Wall clearance back [mm] | 100 | 100 | 100 | 100 |
Wall clearance sidewise [mm] | 175 | 175 | 175 | 175 |
Measures - Doors | ||||
Unit doors | 1 | 1 | 1 | 1 |
Measures - Internal Dimensions | ||||
Width [mm] | 619 | 619 | 619 | 619 |
Height [mm] | 1300 | 1300 | 1300 | 1300 |
Depth [mm] | 600 | 600 | 600 | 600 |
Measures | ||||
Interior volume [L] | 483 | 483 | 483 | 483 |
Net weight of the unit (empty) [kg] | 320 | 320 | 320 | 320 |
permitted load [kg] | 200 | 200 | 200 | 200 |
Permitted load of individual compartment [kg] | 50 | 50 | 50 | 50 |
Load per rack [kg] | 50 | 50 | 50 | 50 |
Environment-specific data | ||||
Energy consumption at set-point -80 °C and ambient temperature = 25 °C (+/- 10 %) [kWh/d] | 13.9 | 13.9 | 13.9 | 13.9 |
Sound-pressure level [dB(A)] | 49 | 49 | 49 | 49 |
Fixtures | ||||
Number of shelves (std./max.) | 3/7 | 3/7 | 3/7 | 3/7 |
Load-variants | ||||
Freezer racks per compartment | 4 | 4 | 4 | 4 |
Cryoboxes, 50 mm | 352 | 352 | 352 | 352 |
* All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.