Category: HPLC, FPLC, & LC/MS | Subcategory: Detectors
Thermo Electron Corporation Accela PDA Detector
ITEM: 27229
This product is currently out of stock, please check our complete line of in-stock Detectors by clicking the green link below or give one of the members of our sales team a call today at (860) 691-2213 so we can help you find a suitable replacement.
The Thermo Electron Corporation Accela PDA Detector P/N 60057-60040 uses thermos scoentofoc light pipe technology to ensure fast photo diode array detection. Thermo Electron Corporation Accela PDA Detector is optimized for the detection of high speed chromatographic separations.
Specifications:
-Wavelength Range: 190–800 nm at 1 nm increment
-Wavelength Accuracy: ±1 nm at 254 nm and 640 nm
-Wavelength Calibration: Using Holmium oxide solution
-Wavelength Resolution: 1.2 nm (512 pixel array)
-Absorbance Non-Linearity: < 5% at 2.0 AU at 257 nm -
-Absorbance Range: -2.0 to +4.0 AU, 20-bit resolution
-Light Source: Pre-aligned Deuterium and Tungsten lamps
-Diodes: 512
-Diode Spacing: 1.2 nm
-Operating Temperature: 10–30 °C
-Ambient Environment: 10–40 °C 5–95% relative humidity (non-condensing)
-Dimensions: 18 ×36 ×47 cm (H ×W ×D)
-Weight: 19.5 kg
-Line voltage: 100/115 or 230 VAC
-Line frequency: 50/60 Hz
-Power consumption: 225 VA max
-Wavelength Accuracy: ±1 nm at 254 nm and 640 nm
-Wavelength Calibration: Using Holmium oxide solution
-Wavelength Resolution: 1.2 nm (512 pixel array)
-Absorbance Non-Linearity: < 5% at 2.0 AU at 257 nm -
-Absorbance Range: -2.0 to +4.0 AU, 20-bit resolution
-Light Source: Pre-aligned Deuterium and Tungsten lamps
-Diodes: 512
-Diode Spacing: 1.2 nm
-Operating Temperature: 10–30 °C
-Ambient Environment: 10–40 °C 5–95% relative humidity (non-condensing)
-Dimensions: 18 ×36 ×47 cm (H ×W ×D)
-Weight: 19.5 kg
-Line voltage: 100/115 or 230 VAC
-Line frequency: 50/60 Hz
-Power consumption: 225 VA max